Mindray bc-3000 Plus 5 DIFF

Description

  • 3-part differentiation of WBC ; 19parameters + 3histograms
  • 2 countingmodes : whole blood and prediluted
  • Throughput : 60 samples per hour
  • Automatic diluting, lyzing, mixing, rinsing and unclogging
  • Automatically cleaning sampling probe
  • Up to 35000 sample results (including histograms) can be stored
  • Built-in thermal recorder, optional external printer
  • Barcode scanner optional
  • Large color LCD display

Technical Specifications

Parameters
WBC, Lymph #, Mid # , Gran # , Lymph# , Mid % , Gran % , Lymph % , RBC , HGB , HCT , MCV , MCH , MCHC , RDW-CV , RDW-SD , PLT , MPV , PDW , PCT and Histogram for WBC , RBCC , PLT

Principles
Electrical impedance method for counting and cyanide free method for hemoglobin

Performance
Parameter Linearity Range Precision ( CV%)
WBC ( 109 / L ) 0.3-99.9 2.5 (7.0-15.0)
RBC ( 1012 / L ) 0.20 – 8.00 2 (3.50 – 6.00)
HGB ( g / L ) 10-250 1.5 (110-180)
MCV (fL) 0.5 (80.0-110.0)
PLT ( 109 / L ) 10-999 5 ( 150-500)

Sample Volume
Prediluted 20 µL
Whole Blood 13 µL

Aperture Diameter
WBC 100 µL
RBC/PLT 70 µL

Oilution Rations
Whole Blood Capillary Blood
WBC / HGB 1 : 308 1 : 428
RBC / PLT 1:44872 1 : 43355

Throughput
60 Samples per hour

Display
Large color LCD display resolution : 800 x 600

Alarms
41 error messages

Carryover
WBC, RBC, HGB <= 0.5%, plt <= 1%

Input Output
RS2322 x 2.1 Parallel printer (optional ), 1 barcode scanner (optional)
1 Keyboard interface

Printout
Thermal recorde, 50 mm wide paper, multitude of printout formats, printer optional

Operating Environment
Temperature 15’’C – 30”C
Humidity 30%-85%

Power Requirement
AC 100-240V, 50/60 Hz

Dimension
390 (W) x 4600 (H) x 400(D)

Weight
21 Kg