Description
Principles | Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method. independent Basophil channel Impedance method for WBC, RBC, PLT counting Cyanide-free reagent for Hemoglobin test |
Parameters | 27 : WBC, Lym%, Mon%, Neu%, Eos, Bas%, Lym#, Mon#, Neu#, Eos#, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, LIC%, LIC#, ALY%, ALY# 3 histograms and 1 scattergram |
Throughput | Up to 60 samples per hour |
Sample Volume | Whole BlooD : 20 µL Prediluted 20 µL |
Test Mode | CBC CBC+DIFF |
Performance | Carryover Precision Linearity WBC < 0.5% < 2.0% (4-15 x 109/L) 0.00-99.99 x109 /L RBC < 0.5% <1.5% (3.5-6.0 x 1012/L) 0.00-8.00 x1012/L HGB < 0.5% <1.5%,(110-180g/L) 0-250g PLT < 1.0% <4.0% (150-500 x 109/L) 0-1000 x109 /L |
Data Storage Capacity | Up to 40,000 results including numeric and graphical information |
Communication | LAN Port supports H17 protocol |
Operating Environment Temperature | 15°C-30°C Humidity: 30-85% Air Pressure: 70-106 kPa |
Power Requirement | .0 100-240V s 300VA 50/601 |
Dimension and Weight | 410mm(L), x 470mm(W), x 530mm(H), Weight<50 Kg |